TM 55-4920-328-13
Figure 5-2. Marking Oil Quantity Probe
CAUTION
Do
not
adjust
the
ZERO
potentiometer at any time. This
adjustment
Is
accomplished
at
manufacture of the unit and is not to
be adjusted at any subsequent time.
Failure to observe this precaution
will result In erroneous readings
displayed
on
the
OIL/FLOW
CONSUMPTION monitor unit.
m. Loosen the locknut on the OIL PROBE DRY
potentiometer (34) and adjust the potentiometer for a
reading of 67.5 + 0.1 LBS on the OIL/FLOW
CONSUMPTION monitor. Tighten the locknut.
n. Using the same lubrication oil as in the system, at
the same temperature as the system, fill the beaker to
the pencil mark on the oil quantity probe.
o. Loosen the locknut on the OIL PROBE FULL
potentiometer (33) and adjust the potentiometer for a
reading of 147.7 + 0.5 LBS on the OIL/FLOW
CONSUMPTION monitor. Tighten the locknut.
p. Repeat steps (m) through (o) as required to
compensate for interaction of adjustments.
q. Remove the oil quantity probe from the beaker,
empty the oil from the beaker, and reinstall the probe in
the lubricating oil unit (6, figure 1-12).
5-18. Checkout of Mechanical Systems.
Checkout
of
METS
mechanical
systems
is
accomplished on a daily basis or prior to use; individual
component checkout is not required. Refer to chapter 2,
Preliminary Checkout Prior to Test paragraph for
information on checkout of mechanical systems.
5-10