TM    55-4920-428-13
TERMINA
L
TB4-10
TB4-9  (record)
TB4-8
TB4-11
TB4-12
TB4-6
MULTIMETER 
INDICATION
(dc volts)
27 to 29
19 to 22
-19  to  -22
11.5 to 12.5
-11.5  to  -12.5
4.7 to 5.3
e. Reference Voltage Test.
(1) Set the multi meter to measure ac volts.
(2) Connect the multi meter to terminals TB4-7 (line) and TB4-1. The multimeter shall indicate
12.9 to 13.1 volts ac.
(3) Connect  the  frequency  counter  to  terminals  TB4-7  (line)  and  TB4-1.  The  frequency  counter
shall indicate 395 to 402 Hz.
f.  DVM  Test.
(1) Set the SYSTEM SELECT switch to BITE.
(2) Set the BITE switch to +20V. The DVM shall indicate within 2% the voltage recorded at ter-
minal TB4-9. (Ref step  d.)
g.  Concluding  Procedure.
(1)
(2)
(3)
4 - 4 1
.
Set the power switch to OFF.
Disconnect  the  power  cable.
Install the panel. (Ref para 4-41.)
Installation    of    Panel.
Close the rear panel. Secure it with three screws.
Install the panel in the case. Use the handles.
Tighten  the  10  captive  screws  to  secure  the  panel  to  the  case.
a. Ensure the power switch is at OFF and the power cable is disconnected.
b.
c.
d.
4-42. Final Test. (CRC)
Test  the  electrical  unit  whenever  performance  is  suspect  and  whenever  the  unit  is  repaired.  Perform  the
final test in the sequence in which the individual tests are presented. To check operation of a particular
circuit, perform the preliminary procedures (para 4-44); then perform the selected circuit test (para 4-
45 through 4-53). Where required, additional conditions are included in notes at the beginning of the
circuit  test.
4-15